Product Name: Acutronic Monsoon III Jet Ventilator - Disinfecting & Cleaning

 


Wipe surfaces of housing, cable and connecting hoses with a cloth soaked with water and a detergent after each patient.

Recommendation:
Change and Sterilize patient connecting tubes with each case to include Jet tube, Airway Tube and Auxiliary flow tube if used to avoid cross contamination.

The following procedures can be used for sterilization:

 



Recommended Sterilization Procedures for patient tubing
Category
Description
Sterilization
Patient Tubing

Patient Connecting
Tube

P/N: ACU-7720

Autoclave: Steam 18 minutes (134)(C)

Gas: Ethylene Oxid

Note: After a maximum of 20 sterilizations the tubes must be replaced and are no longer usable!

Patient Airway Tubing

Airway Proximal Monitor Tube

P/N: ACU-7723

Autoclave: Steam 18 minutes (134)(C)

Gas: Ethylene Oxid

Note: After a maximum of 20 sterilizations the tubes must be replaced and are no longer usable!

Bypass Flow Tubing

Bypass Flow Tube (AUX Gas Tube)

P/N: ACU-7722

Autoclave: Steam 18 minutes (134)(C)

Gas: Ethylene Oxid

Note: After a maximum of 20 sterilizations the tubes must be replaced and are no longer usable!