Product Name: Acutronic Jet Ventilator - Disinfecting & Cleaning

 


Wipe surfaces of housing, cable and connecting hoses with a cloth soaked with water and a detergent after each patient.

Recommendation:
Change and Sterilize patient connecting tubes with each case to include Jet tube, Airway Tube and Auxiliary flow tube if used to avoid cross contamination. The humidifier pump tubing should be changed once a day.

The following procedures can be used for sterilization:

 



Recommended Sterilization Procedures for patient tubing
Category
Description
Sterilization
Patient Tubing

Patient Tube
P/N: ACU-1004

Autoclave: Steam 18 minutes at 15 PSI (134)(C)
Radiation: Radiation up to 2.5 MRad.
Gas: Ethylene Oxid

Note: After a maximum of 20 sterilizations the tubes must be replaced and are no longer usable!

Patient Airway Tubing

Patient Airway
Tube
P/N: ACU-1005

Autoclave: Steam 18 minutes at 15 PSI (134)(C)
Radiation: Radiation up to 2.5 MRad.
Gas: Ethylene Oxid

Note: After a maximum of 20 sterilizations the tubes must be replaced and are no longer usable!

Bypass Flow Tubing

Bypass Flow
Tube
P/N: ACU-1003

Autoclave: Steam 18 minutes at 15 PSI (134)(C)
Radiation: Radiation up to 2.5 MRad.
Gas: Ethylene Oxid

Note: After a maximum of 20 sterilizations the tubes must be replaced and are no longer usable!